HPLC Troubleshooting Guide
SCION Instruments have released their LC6000 Troubleshooting guide. This guide will provide end users with common troubleshooting techniques to identify and resolve any HPLC issues, ensuring minimum downtime and maximum productivity.
Introduction
The following guide will demonstrate troubleshooting to identify and solve your HPLC problems. At any one time, only change one component of the instrument configuration.
It is important to document all changes and results observed when troubleshooting and should be referred to if future issues occur. Always compare changed parameters with previous results to ensure best system performance is obtained. Working with a logical and methodical approach will help isolate and rectify possible causes.
Diagnostic and Problem Solving
Retention Time Drifts
Cause |
Solution |
|
- Use thermostat column oven
- Change temperature of column oven
|
- Incorrect mobile phase composition
|
- Prepare fresh mobile phase
|
- Poor column equilibration
|
- Increase column equilibration time
|
|
- Reset flow rate and test using liquid flow meter
|
|
|
Baseline Drifts
Cause |
Solution |
- Column temperature fluctuation
|
- Use thermostat column oven
- Use a thermometer to check set column temperature is accurate
|
- Incorrect mobile phase composition
|
- Prepare fresh mobile phase
- Check mixer is working for gradient methods
|
- Contamination of detector flow cell
|
- Flush flow cell with a strong organic solvent
- Replace flow cell
|
|
- Check flow cell is not damaged, replace if necessary
|
|
- Replace outlet or remove blockage
|
|
- Reset flow rate and test using flow meter
|
- Poor column equilibration
|
- Increase column equilibration time
- f recently changed mobile phase, purge the
system of old solvent and pump new mobile
through with 20 column volumes
|
- Retained peaks look like baseline drift
|
- Use a guard column
- Thoroughly flush the column with strong
solvent before next injection
|
Diagnostic and Problem Solving
Baseline Drifts
Cause |
Solution |
- UV Detector not set to maximum absorbance
|
- Use maximum absorbance wavelength of target compound(s)
|
- Reference wavelength of Diode Array Detector is set to the same absorbance wavelength of target compound
|
- Ensure reference wavelength of the detector is different to target compounds
|
Baseline Pulsing
Cause |
Solution |
- Proportioning valves are sticking
|
- Clean thoroughly
- Replace values
|
- Debris in the detector flow cell
|
- Look through the flow cell, if contaminated contact SCION engineer for maintenance
|
Extra Peaks
Cause |
Solution |
|
- Flush system with strong solvent
- Use guard column
- Change guard column
- Filter sample
|
|
- Increase run time
- Increase gradient
- Flush system with strong solvent
|
|
- Prepare fresh mobile phase
- Reduce injection volume
|
Peak Tailing
Cause |
Solution |
|
- Use narrower and shorter PEEK tubing
|
- Prolonged retention of target compound(s)
|
- Modify mobile phase composition
- Use appropriate mobile phase buffer
- Use a different stationary phase column
|
|
- Reverse phase flush column with strong solvent
- Replace column
|
Diagnostic and Problem Solving
Peak Tailing
Cause |
Solution |
|
- Reserve phase flush column with strong solvent
- Replace column
|
|
- Change mobile phase composition
- Increase gradient program
- Use longer column
|
|
- Adjust mobile phase pH
- Prepare new mobile phase with correct pH
|
|
|
Peak Fronting
Cause |
Solution |
- Column temperature too low
|
- Increase column temperature
- Use thermostat column oven
|
|
- Reduce injection volume
- Dilute sample concentration
|
- Wrong mobile phase composition
|
- Prepare fresh mobile phase
|
- Column stationary phase depleted
|
- Replace the column
- Use new column with different stationary phase
|