Cannabis Potency Analysis by GC-FID with Nitrogen Carrier Gas

Introduction

Hemp and marijuana are becoming more popular because of legalization in multiple countries. The increase of this market also increases the offer of products that contain very different levels of cannabinoids than the content described on the label. Therefore, as quality control it is important to monitor the level of cannabinoids.

Marijuana usually contains high levels of Δ9- THC and a lower level of CBD, with hemp it is the other way around. The primary psychoactive component is Δ9-THC, while CBD is the primary
therapeutic component.

Consumer hemp generally comes in the form of hemp oil which is used for medical purposes while marijuana is often smoked. Keep in mind that both samples need different sample preparation before analysis due to the difference in sample matrixes.

Potency analysis identifies different cannabinoids and measures their concentration as an indication of the strength of the product. There are over 500 chemical compounds in Cannabis, the most tested cannabinoids can be found in table 1. These are also discussed in this application note.

Component Comment
CBC (Cannabichromene)
CBD (Cannabidiol)
Δ8-THC (Δ8-tetrahydrocannabinol)
Δ9-THC (Δ9-tetrahydrocannabinol)
CBG (Cannabigerol)
CBN (Cannabinol)
CBDV (Cannabidivarin) Not detectable by GC unless the sample is derivatised.
CBGA (Cannabigerolic acid) Not detectable by GC unless the sample is derivatised.
THC-A (Δ9-tetrahydrocannabiolic acid) Not detectable by GC unless the sample is derivatised.
CBDA (Cannabiolic acid) Not detectable by GC unless the sample is derivatised.

Table 1. Potency components

Injector Splitless 10:1, 275 °C
Column SCION-35MS
Oven Program 150°C (0.1 min), 25°C/min to 310°C (1.0 min)
Carrier Nitrogen, 0.6 ml/min
Detector FID with ceramic jet, 300°C
Air: 300 ml/min, Fuel gas (H2): 30 ml/min, Make up (N2): 30 ml/min
Inj. Volume 0.5 µl
Autosampler 8400
Software Compass CDS

Table 2. Analytical conditions

Experimental

This analysis can be implemented on both the 436-GC as well as on the 456-GC platform. The Scion 436-GC analyser used for this application is equipped with an 8400 autosampler. The analysis of these components can be done using a GC with a FID detector, but it is also possible to use a Mass Spectrometer (MS). A different method is using a HPLC, in this case it is also possible to detect THC-A, CBDA, CBDV and CBGA with extra sample preparation
(derivatization).

The focus of the calibration lies on six components: CBC, CBD, Δ8-THC, Δ9-THC, CBG and CBN. Unfortunately, CBDV, CBGA, THC-A and CBDA cannot be detected without extra sample preparation due to decarboxylation caused by the heat of the injector. Decarboxylation is the removal of a carboxyl group releasing carbon dioxide. CBDV decarboxylates to CBD, CBGA to CBG, THC-A to THC and CBDA to CBD.

The Quality Control (QC) sample was made from 30 µg/ml for all components. Different products were examined to determine potency. Cannabis oil and medical marijuana were chosen in this study.

The cannabis oil was diluted with methanol, for the medical marijuana a bit more preparation was necessary. In order to extract the potency from the medical marijuana, it had to be dried for 2 hours, after that it had to be grinded. 30 ml of methanol was added and placed in an ultrasonic bath for 30 minutes. Evaporate to dryness, reconstitute in methanol. When looking at the settings of the GC it is important that the injector temperature is high enough to prevent carry over. Carry over can occur from previous runs or extra material that is being introduced into the system.

The chromatogram in figure 1 shows carry over from the previous run. When the injector temperature was increased it showed that the carry over disappeared (see figure 2). Another solution for this problem could be a washing method between samples.

Both figure 1 and 2 are examples from Scion Instruments application note AN091 which explains the analyses of potency using a helium carrier gas instead of nitrogen.


Figure 1. Chromatogram of a potency standard showing carry over (A) using helium carrier gas (AN0091).


Figure 2. Chromatogram of a potency standard using helium carrier gas (AN0091).

Peak ID Compound Name
1 CBC*
2 CBD*
3 8-THC
4 9-THC
5 CBG
6 CBN
A Carry over peak

*With helium carrier gas 1 is CBD and 2 is CBC, the switch of the components is caused by using a 5 5-ms column instead of a 35MS


Figure 3. Chromatogram of a potency standard using Nitrogen carrier.

Results

The precision of the method was obtained by ten consecutive injections containing 50 µg/ml of each component. Most standards described that the RSD% of an accurate method must be no more than 2%, CBC showed an RSD of 1.4%, CBD 1.9%, Δ8 Δ8-THC 1.2%, Δ9 Δ9-THC 1.1%, CBG 1.4% and CBN 1.1%. The precision performance of this method is excellent.

When looking at the chromatogram of figure 3 it is shown that all components are base line separated, this means accurate quantitation can take place. In addition to these results, it is shown that the peak tailing factor (PTF) of all components is smaller than 2. CBC 1.1, CBD 1.0, 8 8-THC 0.99, 9 9-THC 1.0, CBG 1.1 and CBN 0.99. This overall is an outstanding result.

The calibration curves for the potency standards were prepared between the 2 to 100 µg/ml. Figure 4 shows the calibration range for all the standards and it showed that all the correlation
coefficients (R R2) bigger than 0.999 which is exceptionally good.

Originally the linearity the limit of detection (LOD) and limit of quantitation (LOQ) was calculated, these results can be found in table 3.

Component LOD (µg/ml) LOQ (µg/ml) 
CBC 1.4 4.4
CBD 0.9 2.6
8-THC 1.0 3.0
9-THC 1.2 3.6
CBG 1.6 4.9
CBN 1.1 3.4

Table 3. LOD and LOQ for the different components

The QC sample was analysed using ten injections. The average concentration for CBC was 27.6 µg/ml with an RSD% of 1.52, CBD was 27.7 µg/ml with an RSD% of 1.24. 8 THC was 28.1 µg/ml with an RSD% of 1.63 , 9 THC was 28.3 µg/ml with an RSD% of 1.36, CBG was 27.6 with an RSD% 1.25 and CBN was 28.0 with an RSD% of 1.49.

The analysis of two cannabis samples was performed on an oil and a medical marijuana sample.

According to the label of the oil sample it should contain 100 mg of CBD, other components were not mentioned on the label After analysis it showed that the oil sample contained 65.5 mg CBD, CBC, 8THC and 9 THC were also detected, unfortunately these were below the LOQ, it also shows that CBG and CBN were not present in this sample.

The medical marijuana did not come with an index stating it’s content. Analyses showed it contained 1.7 % CBC, 12.8% CBD and 1.3% 9 THC. 8 THC and CBG were detected but the concentration was below the LOQ,. CBN was not present in this sample.

Figure 6 shows an example chromatogram of the oil and medical marijuana sample.

Peak ID Compound Name
1 CBC
2 CBD
3 8-THC
4 9-THC
5 CBG
6 CBN


Figure 4. Calibration curves of the potency standards


Figure 6. Chromatogram of an oil and medical marijuana sample

The Scion 4X6-GC analyser equipped with a split/spitless injector, Scion Instruments column and FID is capable of analysing potency from cannabis products in a qualitative and quantitative way.

Although the method developed is well suited for potency analysis it is confirmed that one of the disadvantages of this method is that the acidic form of the potency components is converted in the hot injector. It should be kept in mind that the levels of CBD, CBG and 9-THC in this case is a summation of the non-acidic and acidic form. If the acidic form has to be analysed it is advised to perform extra sample preparation or another analysing method in the form of a Scion Instruments 6000 HPLC method can be chosen.


Picture 1: The Scion 4X6-GC series with 8400 autosampler

Download Application Note

Download complete Application Note: Cannabis Potency Analysis by GC-FID with Nitrogen Carrier Gas

Keep in Touch

If you wish to speak to a member of our team about more information, please don’t hesitate to contact us. Or if you wish to keep up to date with SCION Instruments latest research and articles, why not join us on social media and sign up to our newsletters today?