GC Liner Guide
In gas chromatography, the liner serves as a crucial component within the instrument, facilitating the transfer of the sample to the GC column for its subsequent separation and analysis. Choosing the appropriate liner significantly influences the efficiency and accuracy of your chromatographic process. The selection of the most suitable liner depends upon various factors, such as the nature of the sample and the specific injection technique required.
Navigating the various liner configurations can be challenging. Our comprehensive guide offers invaluable insights into the diverse range of liners available from SCION. This resource aims to empower you in selecting the ideal liner tailored to your unique needs.
GC Liner Selection Guide
Type | Dimensions (mm x mm OD x mm ID) | Part Number | Image | Benefits |
---|---|---|---|---|
Straight | 78.5 x 6.3 x 4.0
78.5 x 6.3 x 2.0 78.5 x 6.3 x 1.8 54.0 x 5.0 x 0.5 |
41312105
41312103 41312108 41312123 |
Split/Splitless PTV (54.0 mm length) General purpose liner Ideal for clean samples Varied concentration of samples Promotes higher split flows More suitable for packed columns Narrow bore (ID<= 2mm) improves focussing of gaseous samples (purge and trap or headspace) 4 mm ID for large sample volume |
|
Straight Quartz Wool | 78.5 x 6.3 x 4.0 | 41312100 | Same benefits as Straight with additional – Helps in evaporation, mixing of analytes Traps non-volatile material Limits discrimination |
|
Straight Quartz Wool Ultra Inert | 78.5 x 6.3 x 4.0 | 41312114 | Same benefits as Straight with additional – Removes possible active sites distributed within the liner’s interior Reduced surface activity Consistent sample vaporization |
|
Focus | 78.5 x 6.3 x 4.0
78.5 x 6.3 x 2.0 |
41312101
41312104 |
Split General purpose Dirty and concentrated samples Contains Quartz Wool and it is Ultra Inert Quartz wool remains in optimum positions due to the necks around Excellent reproducibility due to the wiping of the needle by the Quartz Wool Minimize discrimination Removes possible active sites distributed within the liner’s interior Reduced surface activity Consistent sample vaporization |
|
Taper Focus | 78.5 x 6.3 x 4.0 | 41312102 | Split/Splitless The same benefits as Focus with additional – Top taper helps in mitigating sample flashback |
|
Double Taper Focus | 54.0 x 5.0 x 3.4 | 41312124 | PTV The same benefits as Taper Focus with additional – Bottom taper directs the sample toward the head of the column Low boiling and active compounds |
|
Taper Fast Focus | 78.5 x 6.3 x 2.3 | 41312112 | Same benefits as Taper Focus with additional – Narrow 2.3 mm ID for use with 0.1 mm ID capillary columns Culminates in smaller peak widths |
|
Taper Focus Ultra | 78.5 x 6.3 x 4.0 | 41312115 | Split/Splitless Same benefits as Taper Focus with additional – Trace level analysis Minimize greatly contaminations due to the reduction of active sites during volatilization |
|
Taper | 78.5 x 6.3 x 4.0 | 41312109 | Split/Splitless Active samples Trace level analysis Top taper helps in mitigating sample flashback Increase peak area |
|
Taper Quartz Wool | 78.5 x 6.3 x 4.0 | 41312111 | Extra – Run dirty samples | |
Taper Quartz Wool Ultra Inert | 78.5 x 6.3 x 4.0 | 41312117 | Extra – Removes possible active sites distributed within the liner’s interior Reduced surface activity Consistent sample vaporization |
|
DBL Taper | 78.5 x 6.3 x 4.0 | 41312110 | Splitless Low boiling point compounds Bottom taper minimizes contact with metal parts of the inlet and focuses sample onto the head of the column Top taper aids in minimizing flashback |
|
Recessed Gooseneck | 78.5 x 6.3 x 2.0 | 41312107 | Splitless Trace analysis of active compounds Great recovery and linearity Less restriction – liner volume |
|
Recessed Gooseneck Quartz Wool | 78.5 x 6.3 x 4.0 | 41312106 | Extra – Helps in evaporation, mixing of analytes Traps non-volatile material Limits discrimination |
|
Cold On-Column (COC) Injector | R 0.25 54.0 x 4.6 x 0.5
R 0.25 54.0 x 4.6 x 0.8 R o.50 54.0 x 4.6 x 0.8 |
41312120
41312121 41312122 |
Direct injection Trace level analysis Aqueous samples Active compounds |
|
S/SL Focus | 54.0 x 5.0 x 3.4 | 41312125 | Same benefits as Focus Optimized for S/SL injector |
|
S/SL Taper | 54.0 x 5.0 x 3.4
54.0 x 5.0 x 2.0 |
41312126
41312127 |
Same benefits as Taper Optimized for S/SL injector Narrow bore (ID <= 2 mm) - Trace level analysis |
|
SPME straight | 78.5 x 6.3 x 0.75
54.0 x 5.0 x 0.80 |
41312128
41312113 |
Splitless PTV (54.0 mm length) Optimized for the Solid Phase MicroExtraction (SPME) sample preparation technique |
For further information/assistance, please get in touch and a member of our team will be happy to help.